Figures of merit (Limits of Dectection – LOD and quantification – LOQ) for the ICP-MS analysis of lanthanides.To determine LOD, five calibration solutions were analyzed in the range 0.1 – 100 µg/L. The corresponding linear calibration equations were derived and LOD calculated from the equation: LOD = (standard deviation of the intercept at the origin / slope) * 3.3. Finally, LOQs were estimated as three times of the corresponding LODs. Column A: Element symbol.Column B: Element isotope used for ICP-MS analysis.Column C: LOD, limit of detection in µg/L.Column D: LOQ, limit of quantification in µg/L.Column E: Accuracy expressed as percentage recovery of each element relative to the corresponding abundance in certified reference material BCR-668.