SITCOMTN-148: LSST Camera Electro-Optical Test (Run 7) Results

View Dataset
LSST Camera Team, ;Antilogus, Pierre;Astier, Pierre;Banovetz, John;Bregeon, Johan;Chiang, James;Digel, Seth W.;Fanning, Kevin;Faris, Yassine;Guillemin, Thibault;Johnson, Anthony S.;Juramy-Gilles, Claire;Lage, Craig S.;Liang, Shuang;MacBride, Sean Patrick;Marshall, Stuart;Neal, Homer;Polin, Daniel;Rasmussen, Andrew P.;Roodman, Aaron;Rykoff, Eli S.;Stalder, Brian;Thayer, John Gregg;Tyson, J. Anthony;Park, Hye Yun;Utsumi, Yousuke;Zhang, Zhuoqi

Description

This note collects results from the LSST Camera electro-optical testing prior to installation on the TMA. We describe the focal plane characterization, optimization, and the resulting default settings. Results from eo-pipe are shown for standard runs such as B-protocols, Dense and SuperDense PTCs, gain stability, OpSim runs of Darks, and Darks with variable delays. We also describe features such as e2v Persistence, ITL phosphorescence in coffee stains, remnant charge near Serial register following saturated images, vampire pixels, ITL dips, and other sensor features.

Citations (0)

Mentions (0)

Metrics

Dataset Index

0.6

FAIR Score

46%

Citations

1

Mentions

0

Metrics Over Time

Publication Details

DOI

Publisher

NSF-DOE Vera C. Rubin Observatory

Assigned Domain

Subfield

Radiology, Nuclear Medicine and Imaging

Field

Medicine

Domain

Health Sciences

Confidence Score

44%

Source

Scholar Data Model

Keywords

79 ASTRONOMY AND ASTROPHYSICS

Normalization Factors

FT

63.46

CTw

1.00

MTw